CVE-2026-64934 (GCVE-0-2026-64934)

Vulnerability from cvelistv5 – Published: 2026-08-11 21:15 – Updated: 2026-08-12 12:57
VLAI
Title
Mira Hormone Monitor, Mira Android App Reliance on untrusted inputs in a security decision
Summary
The Mira cloud API accepts the firmware version reported by the companion app as authoritative for a given device, without independently attesting the version from the device itself. An authenticated attacker could submit arbitrary firmware version strings for their own device, allowing them to evade vendor-side vulnerable-fleet analytics, suppress security update prompts to the user, and misrepresent patch-adoption metrics.
SSVC
Exploitation: none Automatable: no Technical Impact: partial
CISA Coordinator · CISA-ADP (v2.0.3)
Decision recorded 2026-08-12 12:56 UTC
CWE
  • CWE-807 - Reliance on untrusted inputs in a security decision
Assigner
icscert CNA
Date Public
2026-08-11 17:00
Show details on NVD website

{
  "containers": {
    "adp": [
      {
        "metrics": [
          {
            "other": {
              "content": {
                "id": "CVE-2026-64934",
                "options": [
                  {
                    "Exploitation": "none"
                  },
                  {
                    "Automatable": "no"
                  },
                  {
                    "Technical Impact": "partial"
                  }
                ],
                "role": "CISA Coordinator",
                "timestamp": "2026-08-12T12:56:24.687141Z",
                "version": "2.0.3"
              },
              "type": "ssvc"
            }
          }
        ],
        "providerMetadata": {
          "dateUpdated": "2026-08-12T12:57:16.516Z",
          "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
          "shortName": "CISA-ADP"
        },
        "title": "CISA ADP Vulnrichment"
      }
    ],
    "cna": {
      "affected": [
        {
          "defaultStatus": "unaffected",
          "product": "Mira Firmware",
          "vendor": "Quanovate Tech Inc. (operating as Mira / Mira Care)",
          "versions": [
            {
              "status": "affected",
              "version": "1.7.1.47",
              "versionType": "custom"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "platforms": [
            "Android"
          ],
          "product": "Mira Android App",
          "vendor": "Quanovate Tech Inc. (operating as Mira / Mira Care)",
          "versions": [
            {
              "status": "affected",
              "version": "4.5.15.4",
              "versionType": "custom"
            }
          ]
        }
      ],
      "credits": [
        {
          "lang": "en",
          "type": "finder",
          "value": "Gigi Xiaoqing Liu, Muzzammil Mohammed, Narmina Karimova, and En Mong of Northeastern University SPQR Lab reported this vulnerability to Quanovate Tech."
        }
      ],
      "datePublic": "2026-08-11T17:00:00.000Z",
      "descriptions": [
        {
          "lang": "en",
          "supportingMedia": [
            {
              "base64": false,
              "type": "text/html",
              "value": "The Mira cloud API accepts the firmware version reported by the companion app as authoritative for a given device, without independently attesting the version from the device itself. An authenticated attacker could submit arbitrary firmware version strings for their own device, allowing them to evade vendor-side vulnerable-fleet analytics, suppress security update prompts to the user, and misrepresent patch-adoption metrics."
            }
          ],
          "value": "The Mira cloud API accepts the firmware version reported by the companion app as authoritative for a given device, without independently attesting the version from the device itself. An authenticated attacker could submit arbitrary firmware version strings for their own device, allowing them to evade vendor-side vulnerable-fleet analytics, suppress security update prompts to the user, and misrepresent patch-adoption metrics."
        }
      ],
      "metrics": [
        {
          "cvssV4_0": {
            "Automatable": "NOT_DEFINED",
            "Recovery": "NOT_DEFINED",
            "Safety": "NOT_DEFINED",
            "attackComplexity": "LOW",
            "attackRequirements": "NONE",
            "attackVector": "NETWORK",
            "baseScore": 5.3,
            "baseSeverity": "MEDIUM",
            "exploitMaturity": "NOT_DEFINED",
            "privilegesRequired": "LOW",
            "providerUrgency": "NOT_DEFINED",
            "subAvailabilityImpact": "NONE",
            "subConfidentialityImpact": "NONE",
            "subIntegrityImpact": "NONE",
            "userInteraction": "NONE",
            "valueDensity": "NOT_DEFINED",
            "vectorString": "CVSS:4.0/AV:N/AC:L/AT:N/PR:L/UI:N/VC:L/VI:N/VA:N/SC:N/SI:N/SA:N",
            "version": "4.0",
            "vulnAvailabilityImpact": "NONE",
            "vulnConfidentialityImpact": "LOW",
            "vulnIntegrityImpact": "NONE",
            "vulnerabilityResponseEffort": "NOT_DEFINED"
          },
          "format": "CVSS",
          "scenarios": [
            {
              "lang": "en",
              "value": "GENERAL"
            }
          ]
        },
        {
          "cvssV3_1": {
            "attackComplexity": "LOW",
            "attackVector": "NETWORK",
            "availabilityImpact": "NONE",
            "baseScore": 4.3,
            "baseSeverity": "MEDIUM",
            "confidentialityImpact": "LOW",
            "integrityImpact": "NONE",
            "privilegesRequired": "LOW",
            "scope": "UNCHANGED",
            "userInteraction": "NONE",
            "vectorString": "CVSS:3.1/AV:N/AC:L/PR:L/UI:N/S:U/C:L/I:N/A:N",
            "version": "3.1"
          },
          "format": "CVSS",
          "scenarios": [
            {
              "lang": "en",
              "value": "GENERAL"
            }
          ]
        }
      ],
      "problemTypes": [
        {
          "descriptions": [
            {
              "cweId": "CWE-807",
              "description": "CWE-807 Reliance on untrusted inputs in a security decision",
              "lang": "en",
              "type": "CWE"
            }
          ]
        }
      ],
      "providerMetadata": {
        "dateUpdated": "2026-08-11T21:15:13.291Z",
        "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "shortName": "icscert"
      },
      "references": [
        {
          "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-26-223-01"
        },
        {
          "url": "https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2026/icsma-26-223-01.json"
        }
      ],
      "solutions": [
        {
          "lang": "en",
          "supportingMedia": [
            {
              "base64": false,
              "type": "text/html",
              "value": "Users should update the Mira app to the latest version iOS v3.5.18 / Android v4.5.18. Firmware v01.07.01.53 is updated via the app when the device is connected. No additional action is required."
            }
          ],
          "value": "Users should update the Mira app to the latest version iOS v3.5.18 / Android v4.5.18. Firmware v01.07.01.53 is updated via the app when the device is connected. No additional action is required."
        }
      ],
      "source": {
        "advisory": "ICSMA-26-223-01",
        "discovery": "EXTERNAL"
      },
      "title": "Mira Hormone Monitor, Mira Android App Reliance on untrusted inputs in a security decision",
      "x_generator": {
        "engine": "Vulnogram 1.0.4"
      }
    }
  },
  "cveMetadata": {
    "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
    "assignerShortName": "icscert",
    "cveId": "CVE-2026-64934",
    "datePublished": "2026-08-11T21:15:13.291Z",
    "dateReserved": "2026-08-03T16:54:56.490Z",
    "dateUpdated": "2026-08-12T12:57:16.516Z",
    "state": "PUBLISHED"
  },
  "dataType": "CVE_RECORD",
  "dataVersion": "5.2",
  "vulnerability-lookup:meta": {
    "vulnrichment": {
      "containers": {
        "adp": [
          {
            "metrics": [
              {
                "other": {
                  "content": {
                    "id": "CVE-2026-64934",
                    "options": [
                      {
                        "Exploitation": "none"
                      },
                      {
                        "Automatable": "no"
                      },
                      {
                        "Technical Impact": "partial"
                      }
                    ],
                    "role": "CISA Coordinator",
                    "timestamp": "2026-08-12T12:56:24.687141Z",
                    "version": "2.0.3"
                  },
                  "type": "ssvc"
                }
              }
            ],
            "providerMetadata": {
              "dateUpdated": "2026-08-12T12:57:13.303Z",
              "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
              "shortName": "CISA-ADP"
            },
            "title": "CISA ADP Vulnrichment"
          }
        ],
        "cna": {
          "affected": [
            {
              "defaultStatus": "unaffected",
              "product": "Mira Firmware",
              "vendor": "Quanovate Tech Inc. (operating as Mira / Mira Care)",
              "versions": [
                {
                  "status": "affected",
                  "version": "1.7.1.47",
                  "versionType": "custom"
                }
              ]
            },
            {
              "defaultStatus": "unaffected",
              "platforms": [
                "Android"
              ],
              "product": "Mira Android App",
              "vendor": "Quanovate Tech Inc. (operating as Mira / Mira Care)",
              "versions": [
                {
                  "status": "affected",
                  "version": "4.5.15.4",
                  "versionType": "custom"
                }
              ]
            }
          ],
          "credits": [
            {
              "lang": "en",
              "type": "finder",
              "value": "Gigi Xiaoqing Liu, Muzzammil Mohammed, Narmina Karimova, and En Mong of Northeastern University SPQR Lab reported this vulnerability to Quanovate Tech."
            }
          ],
          "datePublic": "2026-08-11T17:00:00.000Z",
          "descriptions": [
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "The Mira cloud API accepts the firmware version reported by the companion app as authoritative for a given device, without independently attesting the version from the device itself. An authenticated attacker could submit arbitrary firmware version strings for their own device, allowing them to evade vendor-side vulnerable-fleet analytics, suppress security update prompts to the user, and misrepresent patch-adoption metrics."
                }
              ],
              "value": "The Mira cloud API accepts the firmware version reported by the companion app as authoritative for a given device, without independently attesting the version from the device itself. An authenticated attacker could submit arbitrary firmware version strings for their own device, allowing them to evade vendor-side vulnerable-fleet analytics, suppress security update prompts to the user, and misrepresent patch-adoption metrics."
            }
          ],
          "metrics": [
            {
              "cvssV4_0": {
                "Automatable": "NOT_DEFINED",
                "Recovery": "NOT_DEFINED",
                "Safety": "NOT_DEFINED",
                "attackComplexity": "LOW",
                "attackRequirements": "NONE",
                "attackVector": "NETWORK",
                "baseScore": 5.3,
                "baseSeverity": "MEDIUM",
                "exploitMaturity": "NOT_DEFINED",
                "privilegesRequired": "LOW",
                "providerUrgency": "NOT_DEFINED",
                "subAvailabilityImpact": "NONE",
                "subConfidentialityImpact": "NONE",
                "subIntegrityImpact": "NONE",
                "userInteraction": "NONE",
                "valueDensity": "NOT_DEFINED",
                "vectorString": "CVSS:4.0/AV:N/AC:L/AT:N/PR:L/UI:N/VC:L/VI:N/VA:N/SC:N/SI:N/SA:N",
                "version": "4.0",
                "vulnAvailabilityImpact": "NONE",
                "vulnConfidentialityImpact": "LOW",
                "vulnIntegrityImpact": "NONE",
                "vulnerabilityResponseEffort": "NOT_DEFINED"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            },
            {
              "cvssV3_1": {
                "attackComplexity": "LOW",
                "attackVector": "NETWORK",
                "availabilityImpact": "NONE",
                "baseScore": 4.3,
                "baseSeverity": "MEDIUM",
                "confidentialityImpact": "LOW",
                "integrityImpact": "NONE",
                "privilegesRequired": "LOW",
                "scope": "UNCHANGED",
                "userInteraction": "NONE",
                "vectorString": "CVSS:3.1/AV:N/AC:L/PR:L/UI:N/S:U/C:L/I:N/A:N",
                "version": "3.1"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-807",
                  "description": "CWE-807 Reliance on untrusted inputs in a security decision",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2026-08-11T21:15:13.291Z",
            "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
            "shortName": "icscert"
          },
          "references": [
            {
              "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-26-223-01"
            },
            {
              "url": "https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2026/icsma-26-223-01.json"
            }
          ],
          "solutions": [
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "Users should update the Mira app to the latest version iOS v3.5.18 / Android v4.5.18. Firmware v01.07.01.53 is updated via the app when the device is connected. No additional action is required."
                }
              ],
              "value": "Users should update the Mira app to the latest version iOS v3.5.18 / Android v4.5.18. Firmware v01.07.01.53 is updated via the app when the device is connected. No additional action is required."
            }
          ],
          "source": {
            "advisory": "ICSMA-26-223-01",
            "discovery": "EXTERNAL"
          },
          "title": "Mira Hormone Monitor, Mira Android App Reliance on untrusted inputs in a security decision",
          "x_generator": {
            "engine": "Vulnogram 1.0.4"
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "assignerShortName": "icscert",
        "cveId": "CVE-2026-64934",
        "datePublished": "2026-08-11T21:15:13.291Z",
        "dateReserved": "2026-08-03T16:54:56.490Z",
        "dateUpdated": "2026-08-12T12:57:16.516Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.2"
    }
  }
}



Log in or create an account to share your comment.




Tags
Taxonomy of the tags.


Loading…

Loading…

Loading…

Drag nodes to arrange them. Double-click a related vulnerability to open it, or a source to visit its link. The Physics button on the left tunes the layout and the simulation; the legend at the bottom left hides or shows nodes by kind and by sighting type.

Forecast uses a logistic model when the trend is rising, or an exponential decay model when the trend is falling. Fitted via linearized least squares.

Sightings

Author Source Type Date Other

Nomenclature

  • Seen: The vulnerability was mentioned, discussed, or observed by the user.
  • Confirmed: The vulnerability has been validated from an analyst's perspective.
  • Published Proof of Concept: A public proof of concept is available for this vulnerability.
  • Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
  • Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
  • Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
  • Not confirmed: The user expressed doubt about the validity of the vulnerability.
  • Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.
Vulnerability timeline

Sightings and recorded dates, grouped by day · UTC

JSON
Events included in the timeline

Loading timeline…


    Loading…

    Loading…

    Loading…

    Related by attack behaviour

    Vulnerabilities whose description is nearest to this one in the vector space of the CIRCL/vulnerability-attack-technique-biencoder model. This is a similarity search over the bi-encoder space (plain cosine), not a classification, and it has no measured accuracy.