CVE-2021-38400 (GCVE-0-2021-38400)
Vulnerability from cvelistv5
Published
2021-10-04 17:35
Modified
2024-09-16 23:31
Severity ?
VLAI Severity ?
EPSS score ?
CWE
- CWE-916 - Use of Password Hash With Insufficient Computational Effort
Summary
An attacker with physical access to Boston Scientific Zoom Latitude Model 3120 can remove the hard disk drive or create a specially crafted USB to extract the password hash for brute force reverse engineering of the system password.
References
| ► | URL | Tags | |||
|---|---|---|---|---|---|
|
|||||
Impacted products
| Vendor | Product | Version | ||
|---|---|---|---|---|
| Boston Scientific | ZOOM LATITUDE |
Version: Model 3120 |
{
"containers": {
"adp": [
{
"providerMetadata": {
"dateUpdated": "2024-08-04T01:37:16.609Z",
"orgId": "af854a3a-2127-422b-91ae-364da2661108",
"shortName": "CVE"
},
"references": [
{
"tags": [
"x_refsource_MISC",
"x_transferred"
],
"url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01"
}
],
"title": "CVE Program Container"
}
],
"cna": {
"affected": [
{
"product": "ZOOM LATITUDE",
"vendor": "Boston Scientific",
"versions": [
{
"status": "affected",
"version": "Model 3120"
}
]
}
],
"credits": [
{
"lang": "en",
"value": "Endres Puschner - Max Planck Institute for Security and Privacy, Bochum, Christoph Saatjohann - FH M\u00fcnster University of Applied Sciences, Christian Dresen - FH M\u00fcnster University of Applied Sciences, and Markus Willing - University of Muenster, discovered these issues as part of broader academic research of cardiac devices and reported them to Boston Scientific."
}
],
"datePublic": "2021-09-30T00:00:00",
"descriptions": [
{
"lang": "en",
"value": "An attacker with physical access to Boston Scientific Zoom Latitude Model 3120 can remove the hard disk drive or create a specially crafted USB to extract the password hash for brute force reverse engineering of the system password."
}
],
"metrics": [
{
"cvssV3_1": {
"attackComplexity": "HIGH",
"attackVector": "PHYSICAL",
"availabilityImpact": "LOW",
"baseScore": 6.9,
"baseSeverity": "MEDIUM",
"confidentialityImpact": "HIGH",
"integrityImpact": "HIGH",
"privilegesRequired": "NONE",
"scope": "CHANGED",
"userInteraction": "REQUIRED",
"vectorString": "CVSS:3.1/AV:P/AC:H/PR:N/UI:R/S:C/C:H/I:H/A:L",
"version": "3.1"
}
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-916",
"description": "CWE-916 Use of Password Hash With Insufficient Computational Effort",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2021-10-04T17:35:28",
"orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"shortName": "icscert"
},
"references": [
{
"tags": [
"x_refsource_MISC"
],
"url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01"
}
],
"source": {
"advisory": "ICSMA-21-273-01",
"defect": [
"CWE-916"
],
"discovery": "EXTERNAL"
},
"title": "Use of Password Hash with Insufficient Computational Effort for Boston Scientific Zoom Latitude",
"workarounds": [
{
"lang": "en",
"value": "Boston Scientific is in the process of transitioning all users to a replacement programmer with enhanced security, the LATITUDE Programming System, Model 3300. Boston Scientific will not issue a product update to address the identified vulnerabilities in the ZOOM LATITUDE Programming System, Model 3120."
}
],
"x_generator": {
"engine": "Vulnogram 0.0.9"
},
"x_legacyV4Record": {
"CVE_data_meta": {
"ASSIGNER": "ics-cert@hq.dhs.gov",
"DATE_PUBLIC": "2021-09-30T21:02:00.000Z",
"ID": "CVE-2021-38400",
"STATE": "PUBLIC",
"TITLE": "Use of Password Hash with Insufficient Computational Effort for Boston Scientific Zoom Latitude"
},
"affects": {
"vendor": {
"vendor_data": [
{
"product": {
"product_data": [
{
"product_name": "ZOOM LATITUDE",
"version": {
"version_data": [
{
"version_value": "Model 3120"
}
]
}
}
]
},
"vendor_name": "Boston Scientific"
}
]
}
},
"credit": [
{
"lang": "eng",
"value": "Endres Puschner - Max Planck Institute for Security and Privacy, Bochum, Christoph Saatjohann - FH M\u00fcnster University of Applied Sciences, Christian Dresen - FH M\u00fcnster University of Applied Sciences, and Markus Willing - University of Muenster, discovered these issues as part of broader academic research of cardiac devices and reported them to Boston Scientific."
}
],
"data_format": "MITRE",
"data_type": "CVE",
"data_version": "4.0",
"description": {
"description_data": [
{
"lang": "eng",
"value": "An attacker with physical access to Boston Scientific Zoom Latitude Model 3120 can remove the hard disk drive or create a specially crafted USB to extract the password hash for brute force reverse engineering of the system password."
}
]
},
"generator": {
"engine": "Vulnogram 0.0.9"
},
"impact": {
"cvss": {
"attackComplexity": "HIGH",
"attackVector": "PHYSICAL",
"availabilityImpact": "LOW",
"baseScore": 6.9,
"baseSeverity": "MEDIUM",
"confidentialityImpact": "HIGH",
"integrityImpact": "HIGH",
"privilegesRequired": "NONE",
"scope": "CHANGED",
"userInteraction": "REQUIRED",
"vectorString": "CVSS:3.1/AV:P/AC:H/PR:N/UI:R/S:C/C:H/I:H/A:L",
"version": "3.1"
}
},
"problemtype": {
"problemtype_data": [
{
"description": [
{
"lang": "eng",
"value": "CWE-916 Use of Password Hash With Insufficient Computational Effort"
}
]
}
]
},
"references": {
"reference_data": [
{
"name": "https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01",
"refsource": "MISC",
"url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01"
}
]
},
"source": {
"advisory": "ICSMA-21-273-01",
"defect": [
"CWE-916"
],
"discovery": "EXTERNAL"
},
"work_around": [
{
"lang": "en",
"value": "Boston Scientific is in the process of transitioning all users to a replacement programmer with enhanced security, the LATITUDE Programming System, Model 3300. Boston Scientific will not issue a product update to address the identified vulnerabilities in the ZOOM LATITUDE Programming System, Model 3120."
}
]
}
}
},
"cveMetadata": {
"assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"assignerShortName": "icscert",
"cveId": "CVE-2021-38400",
"datePublished": "2021-10-04T17:35:28.144047Z",
"dateReserved": "2021-08-10T00:00:00",
"dateUpdated": "2024-09-16T23:31:04.619Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.1"
}
Loading…
Loading…
Sightings
| Author | Source | Type | Date |
|---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or seen somewhere by the user.
- Confirmed: The vulnerability is confirmed from an analyst perspective.
- Exploited: This vulnerability was exploited and seen by the user reporting the sighting.
- Patched: This vulnerability was successfully patched by the user reporting the sighting.
- Not exploited: This vulnerability was not exploited or seen by the user reporting the sighting.
- Not confirmed: The user expresses doubt about the veracity of the vulnerability.
- Not patched: This vulnerability was not successfully patched by the user reporting the sighting.
Loading…
Loading…