CVE-2019-25721 (GCVE-0-2019-25721)
Vulnerability from cvelistv5
Published
2026-06-02 17:29
Modified
2026-06-03 16:06
Severity ?
7.1 (High) - CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N
6.5 (Medium) - CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
6.5 (Medium) - CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
VLAI Severity ?
EPSS score ?
CWE
- CWE-400 - Uncontrolled Resource Consumption
Summary
Dräger Infinity M300 patient worn monitors with software version VG2.3.1 and earlier contain a network-based denial of service vulnerability that allows network-adjacent attackers to repeatedly trigger device reboots by sending malicious requests over the Infinity Network. Attackers can exploit this vulnerability to force the device into a fail state requiring manual restart, causing loss of wireless connectivity and interruption of patient monitoring functionality.
References
Impacted products
| Vendor | Product | Version | ||
|---|---|---|---|---|
| Dräger | Infinity M300 |
Version: 0 < Patch: VG2.3.2 |
{
"containers": {
"adp": [
{
"metrics": [
{
"other": {
"content": {
"id": "CVE-2019-25721",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "no"
},
{
"Technical Impact": "partial"
}
],
"role": "CISA Coordinator",
"timestamp": "2026-06-03T12:52:13.424163Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2026-06-03T12:52:26.546Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"defaultStatus": "unknown",
"product": "Infinity M300",
"vendor": "Dr\u00e4ger",
"versions": [
{
"lessThanOrEqual": "VG2.3.1",
"status": "affected",
"version": "0",
"versionType": "custom"
},
{
"status": "unaffected",
"version": "VG2.3.2",
"versionType": "custom"
}
]
}
],
"datePublic": "2019-10-09T00:00:00.000Z",
"descriptions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "\u003cp\u003eDr\u00e4ger Infinity M300 patient worn monitors with software version VG2.3.1 and earlier contain a network-based denial of service vulnerability that allows network-adjacent attackers to repeatedly trigger device reboots by sending malicious requests over the Infinity Network. Attackers can exploit this vulnerability to force the device into a fail state requiring manual restart, causing loss of wireless connectivity and interruption of patient monitoring functionality.\u003c/p\u003e"
}
],
"value": "Dr\u00e4ger Infinity M300 patient worn monitors with software version VG2.3.1 and earlier contain a network-based denial of service vulnerability that allows network-adjacent attackers to repeatedly trigger device reboots by sending malicious requests over the Infinity Network. Attackers can exploit this vulnerability to force the device into a fail state requiring manual restart, causing loss of wireless connectivity and interruption of patient monitoring functionality."
}
],
"metrics": [
{
"cvssV4_0": {
"Automatable": "NOT_DEFINED",
"Recovery": "NOT_DEFINED",
"Safety": "NOT_DEFINED",
"attackComplexity": "LOW",
"attackRequirements": "NONE",
"attackVector": "ADJACENT",
"baseScore": 7.1,
"baseSeverity": "HIGH",
"exploitMaturity": "NOT_DEFINED",
"privilegesRequired": "NONE",
"providerUrgency": "NOT_DEFINED",
"subAvailabilityImpact": "NONE",
"subConfidentialityImpact": "NONE",
"subIntegrityImpact": "NONE",
"userInteraction": "NONE",
"valueDensity": "NOT_DEFINED",
"vectorString": "CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N",
"version": "4.0",
"vulnAvailabilityImpact": "HIGH",
"vulnConfidentialityImpact": "NONE",
"vulnIntegrityImpact": "NONE",
"vulnerabilityResponseEffort": "NOT_DEFINED"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
},
{
"cvssV3_1": {
"attackComplexity": "LOW",
"attackVector": "ADJACENT_NETWORK",
"availabilityImpact": "HIGH",
"baseScore": 6.5,
"baseSeverity": "MEDIUM",
"confidentialityImpact": "NONE",
"integrityImpact": "NONE",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H",
"version": "3.1"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-400",
"description": "CWE-400 Uncontrolled Resource Consumption",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2026-06-03T16:06:32.703Z",
"orgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
"shortName": "VulnCheck"
},
"references": [
{
"tags": [
"vendor-advisory"
],
"url": "https://static.draeger.com/security/download/2019-09-13-Cybersecurity-M300-VGxx-Security-advisory-update_v1-3.pdf"
},
{
"tags": [
"third-party-advisory"
],
"url": "https://www.vulncheck.com/advisories/dr-ger-infinity-m300-vg2-network-based-denial-of-service"
}
],
"source": {
"discovery": "INTERNAL"
},
"title": "Dr\u00e4ger Infinity M300 VG2.3.1 Network-Based Denial of Service",
"x_generator": {
"engine": "Vulnogram 1.0.2"
}
}
},
"cveMetadata": {
"assignerOrgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
"assignerShortName": "VulnCheck",
"cveId": "CVE-2019-25721",
"datePublished": "2026-06-02T17:29:08.696Z",
"dateReserved": "2026-06-02T17:23:43.821Z",
"dateUpdated": "2026-06-03T16:06:32.703Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.2",
"vulnerability-lookup:meta": {
"vulnrichment": {
"containers": "{\"cna\": {\"title\": \"Dr\\u00e4ger Infinity M300 VG2.3.1 Network-Based Denial of Service\", \"source\": {\"discovery\": \"INTERNAL\"}, \"metrics\": [{\"format\": \"CVSS\", \"cvssV4_0\": {\"Safety\": \"NOT_DEFINED\", \"version\": \"4.0\", \"Recovery\": \"NOT_DEFINED\", \"baseScore\": 7.1, \"Automatable\": \"NOT_DEFINED\", \"attackVector\": \"ADJACENT\", \"baseSeverity\": \"HIGH\", \"valueDensity\": \"NOT_DEFINED\", \"vectorString\": \"CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N\", \"exploitMaturity\": \"NOT_DEFINED\", \"providerUrgency\": \"NOT_DEFINED\", \"userInteraction\": \"NONE\", \"attackComplexity\": \"LOW\", \"attackRequirements\": \"NONE\", \"privilegesRequired\": \"NONE\", \"subIntegrityImpact\": \"NONE\", \"vulnIntegrityImpact\": \"NONE\", \"subAvailabilityImpact\": \"NONE\", \"vulnAvailabilityImpact\": \"HIGH\", \"subConfidentialityImpact\": \"NONE\", \"vulnConfidentialityImpact\": \"NONE\", \"vulnerabilityResponseEffort\": \"NOT_DEFINED\"}, \"scenarios\": [{\"lang\": \"en\", \"value\": \"GENERAL\"}]}, {\"format\": \"CVSS\", \"cvssV3_1\": {\"scope\": \"UNCHANGED\", \"version\": \"3.1\", \"baseScore\": 6.5, \"attackVector\": \"ADJACENT_NETWORK\", \"baseSeverity\": \"MEDIUM\", \"vectorString\": \"CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H\", \"integrityImpact\": \"NONE\", \"userInteraction\": \"NONE\", \"attackComplexity\": \"LOW\", \"availabilityImpact\": \"HIGH\", \"privilegesRequired\": \"NONE\", \"confidentialityImpact\": \"NONE\"}, \"scenarios\": [{\"lang\": \"en\", \"value\": \"GENERAL\"}]}], \"affected\": [{\"vendor\": \"Dr\\u00e4ger\", \"product\": \"Infinity M300\", \"versions\": [{\"status\": \"affected\", \"version\": \"0\", \"versionType\": \"custom\", \"lessThanOrEqual\": \"VG2.3.1\"}, {\"status\": \"unaffected\", \"version\": \"VG2.3.2\", \"versionType\": \"custom\"}], \"defaultStatus\": \"unknown\"}], \"datePublic\": \"2019-10-09T00:00:00.000Z\", \"references\": [{\"url\": \"https://static.draeger.com/security\", \"tags\": [\"vendor-advisory\"]}, {\"url\": \"https://www.vulncheck.com/advisories/dr-ger-infinity-m300-vg2-network-based-denial-of-service\", \"tags\": [\"third-party-advisory\"]}], \"descriptions\": [{\"lang\": \"en\", \"value\": \"Dr\\u00e4ger Infinity M300 patient worn monitors with software version VG2.3.1 and earlier contain a network-based denial of service vulnerability that allows network-adjacent attackers to repeatedly trigger device reboots by sending malicious requests over the Infinity Network. Attackers can exploit this vulnerability to force the device into a fail state requiring manual restart, causing loss of wireless connectivity and interruption of patient monitoring functionality.\", \"supportingMedia\": [{\"type\": \"text/html\", \"value\": \"\u003cp\u003eDr\\u00e4ger Infinity M300 patient worn monitors with software version VG2.3.1 and earlier contain a network-based denial of service vulnerability that allows network-adjacent attackers to repeatedly trigger device reboots by sending malicious requests over the Infinity Network. Attackers can exploit this vulnerability to force the device into a fail state requiring manual restart, causing loss of wireless connectivity and interruption of patient monitoring functionality.\u003c/p\u003e\", \"base64\": false}]}], \"problemTypes\": [{\"descriptions\": [{\"lang\": \"en\", \"type\": \"CWE\", \"cweId\": \"CWE-400\", \"description\": \"CWE-400 Uncontrolled Resource Consumption\"}]}], \"providerMetadata\": {\"orgId\": \"83251b91-4cc7-4094-a5c7-464a1b83ea10\", \"shortName\": \"VulnCheck\", \"dateUpdated\": \"2026-06-02T17:29:46.712Z\"}}, \"adp\": [{\"title\": \"CISA ADP Vulnrichment\", \"metrics\": [{\"other\": {\"type\": \"ssvc\", \"content\": {\"id\": \"CVE-2019-25721\", \"role\": \"CISA Coordinator\", \"options\": [{\"Exploitation\": \"none\"}, {\"Automatable\": \"no\"}, {\"Technical Impact\": \"partial\"}], \"version\": \"2.0.3\", \"timestamp\": \"2026-06-03T12:52:13.424163Z\"}}}], \"providerMetadata\": {\"shortName\": \"CISA-ADP\", \"orgId\": \"134c704f-9b21-4f2e-91b3-4a467353bcc0\", \"dateUpdated\": \"2026-06-03T12:52:20.275Z\"}}]}",
"cveMetadata": "{\"cveId\": \"CVE-2019-25721\", \"state\": \"PUBLISHED\", \"dateUpdated\": \"2026-06-02T17:29:46.712Z\", \"dateReserved\": \"2026-06-02T17:23:43.821Z\", \"assignerOrgId\": \"83251b91-4cc7-4094-a5c7-464a1b83ea10\", \"datePublished\": \"2026-06-02T17:29:08.696Z\", \"assignerShortName\": \"VulnCheck\"}",
"dataType": "CVE_RECORD",
"dataVersion": "5.2"
}
}
}
Loading…
Loading…
Sightings
| Author | Source | Type | Date |
|---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or seen somewhere by the user.
- Confirmed: The vulnerability is confirmed from an analyst perspective.
- Published Proof of Concept: A public proof of concept is available for this vulnerability.
- Exploited: This vulnerability was exploited and seen by the user reporting the sighting.
- Patched: This vulnerability was successfully patched by the user reporting the sighting.
- Not exploited: This vulnerability was not exploited or seen by the user reporting the sighting.
- Not confirmed: The user expresses doubt about the veracity of the vulnerability.
- Not patched: This vulnerability was not successfully patched by the user reporting the sighting.
Loading…
Loading…