Refine your search

1 vulnerability found for Vagus Nerve Stimulator by Pulsetto

CVE-2026-18844 (GCVE-0-2026-18844)
Vulnerability from cvelistv5
Published
2026-08-11 20:03
Modified
2026-08-11 20:13
CWE
Summary
The firmware of the Pulsetto Vagus Nerve Stimulator accepts several undisclosed commands over its Bluetooth Low Energy (BLE) interface. These commands are sent without authentication or encryption, and are never issued by the companion mobile application, yet are fully processed by the device when it is powered on.
Impacted products
Show details on NVD website


{
  "containers": {
    "adp": [
      {
        "metrics": [
          {
            "other": {
              "content": {
                "id": "CVE-2026-18844",
                "options": [
                  {
                    "Exploitation": "none"
                  },
                  {
                    "Automatable": "no"
                  },
                  {
                    "Technical Impact": "partial"
                  }
                ],
                "role": "CISA Coordinator",
                "timestamp": "2026-08-11T20:13:49.240030Z",
                "version": "2.0.3"
              },
              "type": "ssvc"
            }
          }
        ],
        "providerMetadata": {
          "dateUpdated": "2026-08-11T20:13:58.827Z",
          "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
          "shortName": "CISA-ADP"
        },
        "title": "CISA ADP Vulnrichment"
      }
    ],
    "cna": {
      "affected": [
        {
          "defaultStatus": "unaffected",
          "product": "Vagus Nerve Stimulator",
          "vendor": "Pulsetto",
          "versions": [
            {
              "status": "affected",
              "version": "All",
              "versionType": "custom"
            }
          ]
        }
      ],
      "credits": [
        {
          "lang": "en",
          "type": "finder",
          "value": "A.C. Buglione reported this vulnerability to CISA."
        }
      ],
      "datePublic": "2026-08-11T17:00:00.000Z",
      "descriptions": [
        {
          "lang": "en",
          "supportingMedia": [
            {
              "base64": false,
              "type": "text/html",
              "value": "The firmware of the\u0026nbsp;Pulsetto Vagus Nerve Stimulator\u0026nbsp;accepts several undisclosed commands over its Bluetooth Low Energy (BLE) interface. These commands are sent without authentication or encryption, and are never issued by the companion mobile application, yet are fully processed by the device when it is powered on."
            }
          ],
          "value": "The firmware of the\u00a0Pulsetto Vagus Nerve Stimulator\u00a0accepts several undisclosed commands over its Bluetooth Low Energy (BLE) interface. These commands are sent without authentication or encryption, and are never issued by the companion mobile application, yet are fully processed by the device when it is powered on."
        }
      ],
      "metrics": [
        {
          "cvssV4_0": {
            "Automatable": "NOT_DEFINED",
            "Recovery": "NOT_DEFINED",
            "Safety": "NOT_DEFINED",
            "attackComplexity": "LOW",
            "attackRequirements": "NONE",
            "attackVector": "ADJACENT",
            "baseScore": 7.2,
            "baseSeverity": "HIGH",
            "exploitMaturity": "NOT_DEFINED",
            "privilegesRequired": "NONE",
            "providerUrgency": "NOT_DEFINED",
            "subAvailabilityImpact": "NONE",
            "subConfidentialityImpact": "NONE",
            "subIntegrityImpact": "NONE",
            "userInteraction": "NONE",
            "valueDensity": "NOT_DEFINED",
            "vectorString": "CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:H/VA:H/SC:N/SI:N/SA:N",
            "version": "4.0",
            "vulnAvailabilityImpact": "HIGH",
            "vulnConfidentialityImpact": "NONE",
            "vulnIntegrityImpact": "HIGH",
            "vulnerabilityResponseEffort": "NOT_DEFINED"
          },
          "format": "CVSS",
          "scenarios": [
            {
              "lang": "en",
              "value": "GENERAL"
            }
          ]
        },
        {
          "cvssV3_1": {
            "attackComplexity": "LOW",
            "attackVector": "ADJACENT_NETWORK",
            "availabilityImpact": "HIGH",
            "baseScore": 8.1,
            "baseSeverity": "HIGH",
            "confidentialityImpact": "NONE",
            "integrityImpact": "HIGH",
            "privilegesRequired": "NONE",
            "scope": "UNCHANGED",
            "userInteraction": "NONE",
            "vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:H",
            "version": "3.1"
          },
          "format": "CVSS",
          "scenarios": [
            {
              "lang": "en",
              "value": "GENERAL"
            }
          ]
        }
      ],
      "problemTypes": [
        {
          "descriptions": [
            {
              "cweId": "CWE-912",
              "description": "CWE-912",
              "lang": "en",
              "type": "CWE"
            }
          ]
        }
      ],
      "providerMetadata": {
        "dateUpdated": "2026-08-11T20:03:24.225Z",
        "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "shortName": "icscert"
      },
      "references": [
        {
          "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-26-223-02"
        },
        {
          "url": "https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2026/icsma-26-223-02.json"
        }
      ],
      "solutions": [
        {
          "lang": "en",
          "supportingMedia": [
            {
              "base64": false,
              "type": "text/html",
              "value": "Pulsetto has not responded to requests to work with CISA to mitigate this vulnerability. Users are encouraged to reach out directly to Pulsetto for assistance at \u003ca href=\"mailto:info@pulsetto.tech\"\u003einfo@pulsetto.tech\u003c/a\u003e."
            }
          ],
          "value": "Pulsetto has not responded to requests to work with CISA to mitigate this vulnerability. Users are encouraged to reach out directly to Pulsetto for assistance at  info@pulsetto.tech mailto:info@pulsetto.tech ."
        }
      ],
      "source": {
        "advisory": "ICSMA-26-223-02",
        "discovery": "EXTERNAL"
      },
      "title": "Pulsetto Vagus Nerve Stimulator Hidden Functionality",
      "x_generator": {
        "engine": "Vulnogram 1.0.4"
      }
    }
  },
  "cveMetadata": {
    "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
    "assignerShortName": "icscert",
    "cveId": "CVE-2026-18844",
    "datePublished": "2026-08-11T20:03:24.225Z",
    "dateReserved": "2026-08-04T14:51:10.448Z",
    "dateUpdated": "2026-08-11T20:13:58.827Z",
    "state": "PUBLISHED"
  },
  "dataType": "CVE_RECORD",