Refine your search

2 vulnerabilities found for Infinity Kappa by Dräger

CVE-2019-25717 (GCVE-0-2019-25717)
Vulnerability from cvelistv5
Published
2026-06-02 13:42
Modified
2026-06-03 13:46
CWE
  • CWE-538 - Insertion of Sensitive Information into Externally-Accessible File or Directory
Summary
Dräger Infinity Delta, Delta XL, and Kappa patient monitors contain an information disclosure vulnerability that allows unauthenticated network attackers to access log files over a network connection. Attackers can retrieve device internals, location information, and wired network configuration details from the exposed log files.
Impacted products
Vendor Product Version
Dräger Infinity Delta Version: all software versions   <
Create a notification for this product.
   Dräger Infinity Delta XL Version: all software versions   <
Create a notification for this product.
   Dräger Infinity Kappa Version: all software versions   <
Create a notification for this product.
Show details on NVD website


{
  "containers": {
    "adp": [
      {
        "metrics": [
          {
            "other": {
              "content": {
                "id": "CVE-2019-25717",
                "options": [
                  {
                    "Exploitation": "none"
                  },
                  {
                    "Automatable": "no"
                  },
                  {
                    "Technical Impact": "partial"
                  }
                ],
                "role": "CISA Coordinator",
                "timestamp": "2026-06-03T13:45:52.446853Z",
                "version": "2.0.3"
              },
              "type": "ssvc"
            }
          }
        ],
        "providerMetadata": {
          "dateUpdated": "2026-06-03T13:46:03.638Z",
          "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
          "shortName": "CISA-ADP"
        },
        "title": "CISA ADP Vulnrichment"
      }
    ],
    "cna": {
      "affected": [
        {
          "defaultStatus": "unaffected",
          "product": "Infinity Delta",
          "vendor": "Dr\u00e4ger",
          "versions": [
            {
              "status": "affected",
              "version": "all software versions",
              "versionType": "custom"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "Infinity Delta XL",
          "vendor": "Dr\u00e4ger",
          "versions": [
            {
              "status": "affected",
              "version": "all software versions",
              "versionType": "custom"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "Infinity Kappa",
          "vendor": "Dr\u00e4ger",
          "versions": [
            {
              "status": "affected",
              "version": "all software versions",
              "versionType": "custom"
            }
          ]
        }
      ],
      "credits": [
        {
          "lang": "en",
          "type": "finder",
          "value": "Marc Ruef and Rocco Gagliardi, scip AG"
        }
      ],
      "datePublic": "2019-01-22T00:00:00.000Z",
      "descriptions": [
        {
          "lang": "en",
          "supportingMedia": [
            {
              "base64": false,
              "type": "text/html",
              "value": "\u003cp\u003eDr\u00e4ger Infinity Delta, Delta XL, and Kappa patient monitors contain an information disclosure vulnerability that allows unauthenticated network attackers to access log files over a network connection. Attackers can retrieve device internals, location information, and wired network configuration details from the exposed log files.\u003c/p\u003e"
            }
          ],
          "value": "Dr\u00e4ger Infinity Delta, Delta XL, and Kappa patient monitors contain an information disclosure vulnerability that allows unauthenticated network attackers to access log files over a network connection. Attackers can retrieve device internals, location information, and wired network configuration details from the exposed log files."
        }
      ],
      "metrics": [
        {
          "cvssV4_0": {
            "Automatable": "NOT_DEFINED",
            "Recovery": "NOT_DEFINED",
            "Safety": "NOT_DEFINED",
            "attackComplexity": "LOW",
            "attackRequirements": "NONE",
            "attackVector": "ADJACENT",
            "baseScore": 5.3,
            "baseSeverity": "MEDIUM",
            "exploitMaturity": "NOT_DEFINED",
            "privilegesRequired": "NONE",
            "providerUrgency": "NOT_DEFINED",
            "subAvailabilityImpact": "NONE",
            "subConfidentialityImpact": "NONE",
            "subIntegrityImpact": "NONE",
            "userInteraction": "NONE",
            "valueDensity": "NOT_DEFINED",
            "vectorString": "CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:L/VI:N/VA:N/SC:N/SI:N/SA:N",
            "version": "4.0",
            "vulnAvailabilityImpact": "NONE",
            "vulnConfidentialityImpact": "LOW",
            "vulnIntegrityImpact": "NONE",
            "vulnerabilityResponseEffort": "NOT_DEFINED"
          },
          "format": "CVSS",
          "scenarios": [
            {
              "lang": "en",
              "value": "GENERAL"
            }
          ]
        },
        {
          "cvssV3_1": {
            "attackComplexity": "LOW",
            "attackVector": "ADJACENT_NETWORK",
            "availabilityImpact": "NONE",
            "baseScore": 4.3,
            "baseSeverity": "MEDIUM",
            "confidentialityImpact": "LOW",
            "integrityImpact": "NONE",
            "privilegesRequired": "NONE",
            "scope": "UNCHANGED",
            "userInteraction": "NONE",
            "vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:L/I:N/A:N",
            "version": "3.1"
          },
          "format": "CVSS",
          "scenarios": [
            {
              "lang": "en",
              "value": "GENERAL"
            }
          ]
        }
      ],
      "problemTypes": [
        {
          "descriptions": [
            {
              "cweId": "CWE-538",
              "description": "CWE-538 Insertion of Sensitive Information into Externally-Accessible File or Directory",
              "lang": "en",
              "type": "CWE"
            }
          ]
        }
      ],
      "providerMetadata": {
        "dateUpdated": "2026-06-02T13:42:35.033Z",
        "orgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
        "shortName": "VulnCheck"
      },
      "references": [
        {
          "tags": [
            "vendor-advisory"
          ],
          "url": "https://static.draeger.com/security"
        },
        {
          "tags": [
            "third-party-advisory"
          ],
          "url": "https://www.vulncheck.com/advisories/dr-ger-infinity-delta-kappa-patient-monitors-unauthenticated-log-file-disclosure"
        }
      ],
      "source": {
        "discovery": "EXTERNAL"
      },
      "title": "Dr\u00e4ger Infinity Delta/Kappa Patient Monitors Unauthenticated Log File Disclosure"
    }
  },
  "cveMetadata": {
    "assignerOrgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
    "assignerShortName": "VulnCheck",
    "cveId": "CVE-2019-25717",
    "datePublished": "2026-06-02T13:42:35.033Z",
    "dateReserved": "2026-06-01T21:15:41.689Z",
    "dateUpdated": "2026-06-03T13:46:03.638Z",
    "state": "PUBLISHED"
  },
  "dataType": "CVE_RECORD",
  "dataVersion": "5.2"
}

CVE-2019-25716 (GCVE-0-2019-25716)
Vulnerability from cvelistv5
Published
2026-06-01 21:15
Modified
2026-06-03 20:06
CWE
  • CWE-15 - External Control of System or Configuration Setting
Summary
Dräger Infinity Delta, Delta XL, and Kappa patient monitors contain a denial-of-service vulnerability that allows remote attackers to cause the monitor to reboot by sending a malformed network packet. Attackers can repeatedly send malformed network packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity.
Impacted products
Vendor Product Version
Dräger Infinity Delta Version: Infinity Delta   <
Create a notification for this product.
Show details on NVD website


{
  "containers": {
    "adp": [
      {
        "metrics": [
          {
            "other": {
              "content": {
                "id": "CVE-2019-25716",
                "options": [
                  {
                    "Exploitation": "none"
                  },
                  {
                    "Automatable": "no"
                  },
                  {
                    "Technical Impact": "partial"
                  }
                ],
                "role": "CISA Coordinator",
                "timestamp": "2026-06-02T15:09:36.384627Z",
                "version": "2.0.3"
              },
              "type": "ssvc"
            }
          }
        ],
        "providerMetadata": {
          "dateUpdated": "2026-06-02T15:45:56.330Z",
          "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
          "shortName": "CISA-ADP"
        },
        "title": "CISA ADP Vulnrichment"
      }
    ],
    "cna": {
      "affected": [
        {
          "defaultStatus": "unaffected",
          "product": "Infinity Delta",
          "vendor": "Dr\u00e4ger",
          "versions": [
            {
              "status": "affected",
              "version": "Infinity Delta",
              "versionType": "custom"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "Infinity Delta XL",
          "vendor": "Dr\u00e4ger",
          "versions": [
            {
              "status": "affected",
              "version": "Infinity Delta XL",
              "versionType": "custom"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "Infinity Kappa",
          "vendor": "Dr\u00e4ger",
          "versions": [
            {
              "status": "affected",
              "version": "Infinity Kappa",
              "versionType": "custom"
            }
          ]
        }
      ],
      "credits": [
        {
          "lang": "en",
          "type": "finder",
          "value": "Marc Ruef and Rocco Gagliardi, scip AG"
        }
      ],
      "datePublic": "2019-01-22T00:00:00.000Z",
      "descriptions": [
        {
          "lang": "en",
          "supportingMedia": [
            {
              "base64": false,
              "type": "text/html",
              "value": "\u003cp\u003eDr\u00e4ger Infinity Delta, Delta XL, and Kappa patient monitors contain a denial-of-service vulnerability that allows remote attackers to cause the monitor to reboot by sending a malformed network packet. Attackers can repeatedly send malformed network packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity.\u003c/p\u003e"
            }
          ],
          "value": "Dr\u00e4ger Infinity Delta, Delta XL, and Kappa patient monitors contain a denial-of-service vulnerability that allows remote attackers to cause the monitor to reboot by sending a malformed network packet. Attackers can repeatedly send malformed network packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity."
        }
      ],
      "metrics": [
        {
          "cvssV4_0": {
            "Automatable": "NOT_DEFINED",
            "Recovery": "NOT_DEFINED",
            "Safety": "NOT_DEFINED",
            "attackComplexity": "LOW",
            "attackRequirements": "NONE",
            "attackVector": "ADJACENT",
            "baseScore": 7.1,
            "baseSeverity": "HIGH",
            "exploitMaturity": "NOT_DEFINED",
            "privilegesRequired": "NONE",
            "providerUrgency": "NOT_DEFINED",
            "subAvailabilityImpact": "NONE",
            "subConfidentialityImpact": "NONE",
            "subIntegrityImpact": "NONE",
            "userInteraction": "NONE",
            "valueDensity": "NOT_DEFINED",
            "vectorString": "CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N",
            "version": "4.0",
            "vulnAvailabilityImpact": "HIGH",
            "vulnConfidentialityImpact": "NONE",
            "vulnIntegrityImpact": "NONE",
            "vulnerabilityResponseEffort": "NOT_DEFINED"
          },
          "format": "CVSS",
          "scenarios": [
            {
              "lang": "en",
              "value": "GENERAL"
            }
          ]
        },
        {
          "cvssV3_1": {
            "attackComplexity": "LOW",
            "attackVector": "ADJACENT_NETWORK",
            "availabilityImpact": "HIGH",
            "baseScore": 6.5,
            "baseSeverity": "MEDIUM",
            "confidentialityImpact": "NONE",
            "integrityImpact": "NONE",
            "privilegesRequired": "NONE",
            "scope": "UNCHANGED",
            "userInteraction": "NONE",
            "vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H",
            "version": "3.1"
          },
          "format": "CVSS",
          "scenarios": [
            {
              "lang": "en",
              "value": "GENERAL"
            }
          ]
        }
      ],
      "problemTypes": [
        {
          "descriptions": [
            {
              "cweId": "CWE-15",
              "description": "CWE-15 External Control of System or Configuration Setting",
              "lang": "en",
              "type": "CWE"
            }
          ]
        }
      ],
      "providerMetadata": {
        "dateUpdated": "2026-06-03T20:06:47.074Z",
        "orgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
        "shortName": "VulnCheck"
      },
      "references": [
        {
          "tags": [
            "vendor-advisory"
          ],
          "url": "https://static.draeger.com/security/download/2019-01-22-draeger-infinity-delta-vf10-1-security-advisory.pdf"
        },
        {
          "tags": [
            "third-party-advisory"
          ],
          "url": "https://www.vulncheck.com/advisories/dr-ger-infinity-delta-kappa-patient-monitor-dos-via-malformed-network-packet"
        }
      ],
      "source": {
        "discovery": "EXTERNAL"
      },
      "title": "Dr\u00e4ger Infinity Delta/Kappa Patient Monitor DoS via Malformed Network Packet",
      "x_generator": {
        "engine": "Vulnogram 1.0.2"
      }
    }
  },
  "cveMetadata": {
    "assignerOrgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
    "assignerShortName": "VulnCheck",
    "cveId": "CVE-2019-25716",
    "datePublished": "2026-06-01T21:15:07.156Z",
    "dateReserved": "2026-06-01T20:44:47.913Z",
    "dateUpdated": "2026-06-03T20:06:47.074Z",
    "state": "PUBLISHED"
  },
  "dataType": "CVE_RECORD",
  "dataVersion": "5.2"
}